Filtered by vendor Baxter
Subscribe
Search
Total
21 CVE
| CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
|---|---|---|---|---|---|
| CVE-2021-43935 | 1 Baxter | 10 Welch Allyn Connex Cardio, Welch Allyn Diagnostic Cardiology Suite, Welch Allyn Hscribe Holter Analysis System and 7 more | 2022-07-25 | 6.8 MEDIUM | 9.8 CRITICAL |
| The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges. | |||||
| CVE-2020-12024 | 1 Baxter | 4 Em1200, Em1200 Firmware, Em2400 and 1 more | 2021-11-04 | 3.6 LOW | 6.1 MEDIUM |
| Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI. | |||||
| CVE-2020-12032 | 1 Baxter | 4 Em1200, Em1200 Firmware, Em2400 and 1 more | 2021-11-04 | 6.4 MEDIUM | 9.1 CRITICAL |
| Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI. | |||||
| CVE-2020-12048 | 1 Baxter | 2 Phoenix X36, Phoenix X36 Firmware | 2020-07-16 | 5.0 MEDIUM | 7.5 HIGH |
| Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool. | |||||
| CVE-2020-12037 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2020-07-14 | 5.0 MEDIUM | 7.5 HIGH |
| Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | |||||
| CVE-2020-12036 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2020-07-14 | 5.0 MEDIUM | 7.5 HIGH |
| Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | |||||
| CVE-2020-12035 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2020-07-14 | 3.6 LOW | 4.9 MEDIUM |
| Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration. | |||||
| CVE-2020-12040 | 1 Baxter | 2 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware | 2020-07-09 | 5.0 MEDIUM | 9.8 CRITICAL |
| Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack. | |||||
| CVE-2020-12039 | 1 Baxter | 2 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware | 2020-07-09 | 2.1 LOW | 2.4 LOW |
| Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed. | |||||
| CVE-2020-12016 | 1 Baxter | 4 Em1200, Em1200 Firmware, Em2400 and 1 more | 2020-07-08 | 10.0 HIGH | 9.8 CRITICAL |
| Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI. | |||||
| CVE-2020-12008 | 1 Baxter | 4 Em1200, Em1200 Firmware, Em2400 and 1 more | 2020-07-08 | 5.0 MEDIUM | 7.5 HIGH |
| Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI. | |||||
| CVE-2020-12041 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2020-07-08 | 7.5 HIGH | 9.4 CRITICAL |
| The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot. | |||||
| CVE-2020-12047 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2020-07-08 | 7.5 HIGH | 9.8 CRITICAL |
| The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials. | |||||
| CVE-2020-12043 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2020-07-08 | 7.5 HIGH | 9.8 CRITICAL |
| The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted. | |||||
| CVE-2020-12045 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2020-07-08 | 7.5 HIGH | 9.8 CRITICAL |
| The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials. | |||||
| CVE-2020-12020 | 1 Baxter | 4 Em1200, Em1200 Firmware, Em2400 and 1 more | 2020-07-08 | 3.6 LOW | 6.1 MEDIUM |
| Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user. | |||||
| CVE-2020-12012 | 1 Baxter | 4 Em1200, Em1200 Firmware, Em2400 and 1 more | 2020-07-07 | 3.6 LOW | 6.1 MEDIUM |
| Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI. | |||||
| CVE-2014-5434 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 5.0 MEDIUM | 9.8 CRITICAL |
| Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes. | |||||
| CVE-2014-5431 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 4.6 MEDIUM | 6.8 MEDIUM |
| Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes. | |||||
| CVE-2014-5433 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 7.5 HIGH | 9.8 CRITICAL |
| An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes. | |||||
| CVE-2014-5432 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 7.5 HIGH | 9.8 CRITICAL |
| Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes. | |||||