CVE-2020-25187 1 Medtronic 2 Mycarelink Smart Model 25000, Mycarelink Smart Model 25000 Firmware 2020-12-15 10.0 HIGH 9.8 CRITICAL
Medtronic MyCareLink Smart 25000 all versions are vulnerable when an attacker who gains auth runs a debug command, which is sent to the reader causing heap overflow in the MCL Smart Reader stack. A heap overflow allows attacker to remotely execute code on the MCL Smart Reader, could lead to control of device.